Abstract
In this study, the physiological control algorithm using sliding mode control method is implemented to track the reference input signal. The controller is developed using feed‐forward part, reference model, and steady‐state flow estimator. The proposed control method is evaluated using a dynamic heart‐pump interaction model incorporating descriptions of the cardiovascular system – rotary blood pump. The immediate response of the controller to preload as well as afterload was studied. Stability and feasibility of the control system were demonstrated through the tests. The results showed that the present controller, which allows the left ventricular to automatically adjust to the right ventricular output, reduces the risk of suction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.