Abstract
This study evaluated four bridged-ethylene hybrid (BEH) columns containing C18 (130 Å), peptide C18 (300 Å), phenyl, or a mixed-mode charged surface hybrid (CSH C18 ) using a wide range of antisense oligonucleotide therapeutics. The BEH C18 , peptide, and phenyl columns were all capable of providing significant retention of oligonucleotide samples across multiple ion-pairing systems using alkylamines and 1,1,1,3,3,3,-hexafluoroisopropanol (HFIP). The retention of the oligonucleotides varied depending on the choice of alkylamine, with the order of retention being dimethylcyclohexylamine > diisopropylethylamine > triethylamine. The selectivity of these columns for several closely eluting impurities was similar. Although overall the C18 , peptide, and phenyl columns were all found to be capable of analyzing oligonucleotide therapeutics, the phenyl column was found to be the most retentive and the C18 column provided the best peak shape. The CSH C18 column was found to be degraded by the alkylamine-HFIP mobile phase despite the mobile phase being within the pH stability range of the column.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.