Abstract
A pressure-controlled water heat pipe was evaluated as a comparison furnace for Standard Platinum Resistance Thermometers (SPRTs). Its calibration uncertainties in the range from 65/spl deg/C to 157/spl deg/C were estimated. In direct comparison of SPRTs with a calibrated SPRT, the combined uncertainty of the comparison is estimated to be 0.081 mK. On using the heat pipe as a calibration facility, the combined uncertainty of the calibration is estimated to be 3.1 mK.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.