Abstract

A new sub-millikelvin calibration facility for the range 0 \(^{\circ }\)C to 30 \(^{\circ }\)C is described, that allows calibration of customer thermometers, other than standard platinum resistance thermometers, with an uncertainty lower than 1 millikelvin. The improvements with respect to the traditional calibration facility are reported with particular emphasis on the temperature control (better than 0.2 mK), resistance measurement and calibration procedure. The new facility was validated by using 6 standard platinum resistance thermometers and the calibration uncertainty in the range from 0 \(^{\circ }\)C to 30 \(^{\circ }\)C amounted to 0.31 mK–0.35 mK. To demonstrate the potentiality of this facility, two oceanographic thermometers, Sea-Bird Electronics SBE 3 and SBE 35, were calibrated with an expanded uncertainty of 0.8 mK (\(k=2\)).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.