Abstract

The pulsing performance of the positron-beam-pulsing system of the slow positron beamline at Kyoto University research Reactor was tested using electron beams trimmed by 5-mm- or 15-mm-diameter apertures. Electron pulses with full width at half maximum of 143 ps were obtained with a 5-mm-diameter beam which corresponds to the diameter of the brightness-enhanced positron beam. This indicates the pulsing system has a sufficient pulsing performance for positron annihilation lifetime measurements.