Abstract

In this paper, the evaluation of the dosimetric capability of a detector based on a CR-39 solid-state nuclear track detector coupled to a 1 cm thickness of PMMA radiator was made with the aim of understanding the applicability of this technique to personal and environmental neutron dosimetry. The dosemeter has been exposed to monoenergetic and quasi-monoenergetic neutron beams at PTB in Braunschweig, Germany and at Ithemba Laboratories, in Faure, South Africa, with peak energies ranging from 0.565 to 100 MeV. The results showed a response that is almost independent of the neutron energy in the whole energy range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.