Abstract

BackgroundRecent studies have stated the relevance of having new parameters to quantify the position and orientation of the scapula with patients standing upright. Although biplanar radiography can provide 3D reconstructions of the scapula and the spine, it is not yet possible to acquire these images with patients in the same position. MethodsTwo pairs of images were acquired, one for the 3D reconstruction of the spine and ribcage and one for the 3D reconstruction of the scapula. Following 3D reconstructions, scapular alignment was performed in two stages, a coarse alignment based on manual annotations of landmarks on the clavicle and pelvis, and an adjusted alignment. Clinical parameters were computed: protraction, internal rotation, tilt and upward rotation. Reproducibility was assessed on an in vivo dataset of upright biplanar radiographs. Accuracy was assessed using supine cadaveric CT-scans and digitally reconstructed radiographs. FindingsThe mean error was less than 2° for all clinical parameters, and the 95% confidence interval for reproducibility ranged from 2.5° to 5.3°. InterpretationThe confidence intervals were lower than the variability measured between participants for the clinical parameters assessed, which indicates that this method has the potential to detect different patterns in pathological populations.