Abstract

Laparoscopic instrument handles design and dimensions are crucial to determine the configuration of surgeons' hand grip and, therefore, can have a deleterious effect on overall surgical efficiency and surgeons' comfort. The aim of this study is to investigate the impact of laparoscopic handle size and hand surface area on surgical task performance. A single-blind, randomized crossover trial was carried out with 29 novice medical students. Participants performed three simulated tasks in "black box" simulators using two scissor-type handles of different sizes. Surgical performance was assessed by the number of errors and time required to complete each task. Hand anthropometric data were measured using a 3D scanner. Execution time was significantly higher when cutting and suturing tasks were performed with the smaller handle. In addition, hand surface area was positively correlated with peg transfer task time when performed with the standard handle and was correlated with cutting task time in small and standard handle groups. We also found positive correlations between execution time and the number of errors executed by larger-handed participants. Our findings indicate that laparoscopic handle size and hand area influence surgical performance, highlighting the importance of considering hand anthropometry variances in surgical instrument design.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call