Abstract

We have evaluated the performance of a centrifugal analyzer, specifically designed for hemostasis tests, for methods based on clotting time as measured by light scattering, or based on splitting of chromogenic substrates. The results were compared with those obtained with a semiautomated coagulometer (clotting tests) or with a manually operated photometer (chromogenic tests). In general, the results with the centrifugal analyzer were at least as precise as those of the comparative methods, with greater output and ease of operation. Disadvantages of the instrument are its partial incompatibility with some commercial reagents and the relative rigidity of the operational parameters, which hinder its use for research.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.