Abstract
In this work, the monoclonal antibodies (McAbs) to α-fetoprotein (AFP) were immobilized on two different solid phases, i.e., magnetic particles (MP) and coated tubes (CT). Based on this, a MP based chemiluminescence enzyme immunoassay (MP-CLEIA) and a CT based CLEIA (CT-CLEIA) were proposed for the evaluation of AFP in human serum and their analytical merits were studied and compared. By detailed discussion of several performance variants, including the concentration of immobilized McAb, dilution ratio of horseradish peroxidase (HRP) labeled McAb (HRP-McAb), total assay time, substrate volume, chemiluminescent kinetics, and hook effect concentration, the advantages of MP-CLEIA became conspicuously apparent. Moreover, in the presence of MP, the catalytic activity of labeled enzyme was kept to high extent and the stability of immunoreagents was satisfied. Finally, 59 human serum samples were detected by the MP-CLEIA and a good correlation was obtained when comparing the results with that from a commercial electrochemiluminescence immunoassay kit.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.