Abstract

A high-throughput and simple chemiluminescence (CL) enzyme immunoassay (CLEIA) for the determination of progesterone (P) in human serum was developed, with the highly sensitive 4-methoxy-4-(3-phosphatephenyl)-spiro-(1,2-dioxetane-3,2'-adamantane) (AMPPD)-alkaline phosphatase (ALP) system as the CL detection system. The results showed that the indirect immobilization of rabbit anti-progesterone polyclonal antibody (RAPA) through secondary antibody exhibited apparent advantages over direct coating in terms of antibody saving and improvement of the coating stability and uniformity. The direct analysis of P in human serum without extraction was realized by using 8-anilino-1-naphthalenesulphonic acid (ANS) to displace P from its binding proteins. The effect of several relevant parameters of the immunoreaction were examined and optimized. Compared with some commercial progesterone kits, the presented CLEIA has higher sensitivity with detection limitation as low as 0.06 ng/mL. The recoveries were 95.9-101%. The coefficient of variation was <8.4% and 9.9% for intra- and inter-assay precision, respectively. This method has been successfully applied to the evaluation of P in human serum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.