Abstract

The current study aimed to evaluate the suitability of polyester resin as an alternative material to polymethyl methacrylate (PMMA) for computed tomography (CT) dosimetry phantoms using the GEANT4/GATE Monte Carlo simulation platform. Cylindrical phantoms (32 cm diameter) constructed of polyester resin and PMMA were simulated and compared in terms of atomic composition, effective atomic number, electron density, mass density, and photon interaction mechanisms. Weighted CT dose index (CTDIw) values were calculated for each phantom at 80, 110, and 130 kVp tube voltages based on measurements of CTDI100,c and CTDI100,p. Results demonstrated that the physical properties of polyester closely matched those of PMMA, and the polyester phantom displayed equivalent dosimetric behavior to the PMMA phantom at all tube voltages tested. CTDIw values from the polyester phantom were within 1.4 % of the PMMA phantom across all tube voltages. Conversion coefficients were derived to equate polyester CTDIw values to PMMA dose equivalents. This study found that a polyester resin phantom exhibited radiation dosimetry commensurate with the standard PMMA phantom for CT dose assessment. Consequently, polyester resin represents a viable substitute material when PMMA is unavailable for construction of CT dosimetry phantoms.