Abstract

We compared critical flicker frequency (CFF) thresholds obtained using a novel portable device "Beacon" with thresholds from the commercially available Lafayette Flicker Fusion System (Lafayette-FFS) in patients with cirrhosis. One hundred fifty-three participants with chronic liver disease underwent CFF testing using Beacon and Lafayette-FFS with a method-of-limits and/or forced-choice protocol. Beacon demonstrated excellent test-retest reliability (intraclass correlation 0.91-0.97) and good correlation with the Lafayette-FFS values (intraclass correlation 0.77-0.84). Forced-choice CFF were on average 4.1 Hz higher than method-of-limits descending CFFs. Beacon can be self-administered by patients with chronic liver disease and cirrhosis to measure CFF, a validated screening test for minimal hepatic encephalopathy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.