Abstract

An in vitro study of platelet concentrates storage for 5 days was performed in PVC bags. Modification of the original three-day containers were introduced by thickness reduction. The results at fifth day were comparable to those at third day in standard plastic bags. During storage, variations of platelet counts were very slight with a low LDH release. PH was stable with a good maintenance of phase microscope platelet morphology. PCO2 and PO2 measurements showed a satisfactory gas permeability which could explain a limited lactate production. If in vivo studies of transfusion recovery confirm these data, platelet concentrates storage could be extended up to 5 days by such modifications of standard three-day PVC bags.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.