Abstract
pH, Pco2, and Po2 measurements have been made over a three-week period on General Diagnostics Blood GAS Control Ampoules and on a tonometered buffer system. The latter seems adequate as a precision and accuracy control of pH, Pco2, and Po2 measurements in a single solution. The former seems adequate as a precision control of these measurements, but, on the particular batch provided, mean values need to be determined within the customer's laboratory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.