Abstract

The estimation scheme of uncertainty of determination of 1-hydroxypyrene (1-OHP) in urine was developed analysing the main stages of the analytical procedure: (1) preparation of 1-OHP standards, (2) creation of the calibration curve for the high performance liquid chromatography (HPLC) analysis method with the evaluation of recovery, (3) measuring procedure of aliquot of urine, (4) adjusting the pH of aliquot and hydrolysis with enzyme, (5) solid phase extraction, (6) concentration of the extract, (7) injection of the extract to chromatograph and analysing by the HPLC method, (8) calculation of 1-OHP mass from the calibration curve, (9) calculation of 1-OHP concentration in urine. The evaluation of the uncertainty is based on quantification of individual components. Combined uncertainty was calculated using the law of propagation of uncertainties according to the EURACHEM/CITAC guidelines. Level dependence of the uncertainty arises from the calibration curve.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.