Abstract

Skin entrance doses (SEDs) were estimated by carrying out measurements of air kerma from 101 X-ray machines installed in 45 major and selected hospitals in the country by using a silicon detector-based dose Test-O-Meter. 1209 number of air kerma measurements of diagnostic projections for adults have been analysed for seven types of common diagnostic examinations, viz. chest (AP, PA, LAT), lumbar spine (AP, LAT), thoracic spine (AP, LAT), abdomen (AP), pelvis (AP), hip joints (AP) and skull (PA, LAT) for different film-screen combinations. The values of estimated diagnostic reference levels (DRLs) (third quartile values of SEDs) were compared with guidance levels/DRLs of doses published by the IAEA-BSS-Safety Series No. 115, 1996; HPA (NRPB) (2000 and 2005), UK; CRCPD/CDRH (USA), European Commission and other national values. The values of DRLs obtained in this study are comparable with the values published by the IAEA-BSS-115 (1996); HPA (NRPB) (2000 and 2005) UK; EC and CRCPD/CDRH, USA including values obtained in previous studies in India.