Abstract

This article deals with the estimation of doses received by patients undergoing radiological examinations in order to establish dose reference levels (DRLs) in Latvia. Several large hospitals, small hospitals and private practices were selected for patient dose measurements. The measurements were carried out using calibrated thermoluminiscence dosemeters attached to the patient's skin. Exposure parameters and patient's data were recorded. The entrance surface doses (ESDs) to patients undergoing several common X-ray examinations (chest AP/PA, chest LAT, lumbar spine AP/PA, lumbar spine LAT and pelvis) were measured. Data concerning the kV(p) settings, used type of films, focus-film distance and the ESD values were analysed and compared with those recommended by the European Community (EC). Among the different hospitals and private practices, discrepancies in the patient doses and techniques used for the examination were found, where the doses exceeded the EC recommended values owing to a very low kV(p) and a very low sensitivity of the screen film combinations used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.