Abstract

The present study evaluated the organ doses, effective doses and conversion factors from the dose-area product to effective dose in pediatric diagnostic cardiac catheterization performed by in-phantom dosimetry and Monte Carlo simulation. The organ and effective doses in 5-y-olds during diagnostic cardiac catheterizations were evaluated using radiophotoluminescence glass dosemeters implanted into a pediatric anthropomorphic phantom and PCXMC software. The mean effective dose was 3.8 mSv (range: 1.8-7.5 mSv). The conversion factors from the dose-area product to effective dose were 0.9 and 1.6 mSv (Gy cm2)-1 for posteroanterior and lateral fluoroscopy, respectively, and 0.9 and 1.5 mSv (Gy cm2)-1 for posteroanterior and lateral cineangiography, respectively. Effective doses evaluated using the pediatric dosimetry system agreed with those obtained using PCXMC software within 12%. The dose data and conversion factors evaluated may guide the estimation of exposure doses in children undergoing diagnostic cardiac catheterization.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.