Abstract

A rapid, reliable and cost effective analytical procedure for the estimation of ibuprofen in pharmaceutical formulations and human urine samples was developed using transmission Fourier Transform Infrared (FT-IR) spectroscopy. For the determination of ibuprofen, a KBr window with 500μm spacer was used to acquire the FT-IR spectra of standards, pharmaceuticals as well as urine samples. Partial least square (PLS) calibration model was developed based on region from 1807 to 1461cm−1 using ibuprofen standards ranging from 10 to 100μgml−1. The developed model was evaluated by cross-validation to determine standard error of the models such as root mean square error of calibration (RMSEC), root mean square error of cross validation (RMSECV) and root mean square error of prediction (RMSEP). The coefficient of determination (R2) achieved was 0.998 with minimum errors in RMSEC, RMSECV and RMSEP with the value of 1.89%, 1.63% and 4.07%, respectively. The method was successfully applied to urine and pharmaceutical samples and obtained good recovery (98–102%).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.