Abstract
The radiation doses absorbed by major organs of males and females were studied from three types of dental X-ray devices. The absorbed doses from cone-beam computed tomography (CBCT), panoramic and intraoral X-ray machines were in the range of 0.23-1314.85 μGy, and were observed to be high in organs and tissues located in or adjacent to the irradiated area, there were discrepancies in organ doses between male and female. Thyroid, salivary gland, eye lens and brain were the organs that received higher absorbed doses. The organ absorbed doses were considerably lower than the diagnostic reference level for dental radiography in China. The calculated effective radiation doses for males and females were 56.63, 8.15, 2.56μSv and 55.18, 8.99, 2.39μSv, respectively, when using CBCT, the panoramic X-ray machine and intraoral X-ray machine. The effective radiation dose caused by CBCT was much higher than those of panoramic and intraoral X-ray machines.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.