PurposeTo assess doses variabilities in the same abdomen and chest RX exams for adults, to check the need for dose harmonization. To calculate Diagnostic Reference Levels (DRL), mandatory in the European Union, for the Algarve district in Portugal. Our results can be a valuable reference for the Portuguese official determination of DRLs, still in progress. MethodWe considered 4,936 abdomen and 41,320 chest radiographs of adults, covering 7 health centres and 35 radiographers in Algarve. Entrance skin dose (ESD) was calculated for each radiograph and the corresponding uncertainty estimated. Mean doses per centre and per technician, and their uncertainties, were calculated to access dose variabilities. DRLs, set at the 3rd quartile of the total ESD distribution, were determined for a standard patient and for intervals of body mass index (BMI) to study their correlation with patient anatomical variations. Standard quartile errors were estimated. ResultsOur results suggest significant dispersion in applied ESDs among different centres and radiographers. Estimates of DRLs also show small fluctuations across years and an important dependence on BMI intervals. For a standard patient, they are 8.7 ± 0.1 (abdomen) and 0.44 ± 0.01 (chest), while the European DRLs are, respectively, 5.1 and 0.2 (all in mGy). ConclusionsResults suggest that there is room for dose optimization and harmonization with European DRLs, urging a national dose survey and the establishment of official national DRLs. Official DRLs in intervals of BMI would be quite beneficial, to avoid unnecessary dose exposures.