Abstract

The aim of this work was to establish national diagnostic reference levels (DRLs) in Ireland and compare these to existing European DRLs where available. This work surveyed all radiological facilities providing radiography, mammography, and dual-energy x-ray absorptiometry (DXA) services in Ireland. A list of common procedures and clinical tasks was established. A national database of service providers was used to identify the appropriate medical radiological facilities providing these services. These facilities were issued with an online survey. National DRLs were set as the 75th percentile of the distribution of median values obtained. A national median dose was also established. The broad categorisation of equipment type was also considered. Where differences between DRLs established using different detector types were deemed statistically significant, equipment-specific national DRLs were established. National DRLs were established for 12 adult radiography projections. Equipment-specific (computed radiography and digital radiography) adult DRLs were established for four radiography projections. Paediatric DRLs were established for 11 radiography projections, including two based on clinical indications, for a range of paediatric weight categories. National DRLs were established for unilateral two-view mammography and breast tomosynthesis as well as for four DXA clinical indications and projections. All but one Irish DRL figure was found to be below or equal to European data. This work provided a unique opportunity to establish national DRLs based on census data for a range of procedures and clinical tasks across radiography, mammography and DXA and compare these with European levels. This work established national diagnostic reference levels (DRLs) based on census data for a range of procedures and clinical tasks across radiography, mammography and dual-energy x-ray absorptiometry. The establishment of national DRLs is an essential component in the optimisation of patient radiation dose. • Diagnostic reference levels are easily measured quantities intended for use as an aid to optimise patient dose and to identify when levels of patient dose are unusually high. • Data from all medical radiological facilities in Ireland was obtained to establish national diagnostic reference level (DRL) values and national median dose values in radiography, x-ray breast imaging and dual-energy x-ray absorptiometry (DXA) scanning and these were compared to existing European DRLs where available. • National DRL values were established for the first time in breast tomosynthesis, DXA scanning, and paediatric radiography.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.