Abstract

Nanoindentation methods are well suited for probing the mechanical properties of a heterogeneous surface, since the probe size and contact volumes are small and localized. However, the nanoindentation method may introduce errors in the computed mechanical properties when indenting near the interface between two materials having significantly different mechanical properties. Here we examine the case where a soft material is loaded in close proximity to an interface of higher modulus, such as the case when indenting bone near a metallic implant. Results are derived from both an approximate analytical quarter-space solution and a finite element model, and used to estimate the error in indentation-determined elastic modulus as a function of the distance from the apex of contact to the dissimilar interface, for both Berkovich and spherical indenter geometries. Sample data reveal the potential errors in mechanical property determination that can occur when indenting near an interface having higher stiffness, or when characterizing strongly heterogeneous materials. The results suggest that caution should be used when interpreting results in the near-interfacial region.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.