Abstract

Scanning tunneling microscopy (STM) is well known for its high lateral resolution and nondestructive surface profiling capability but its potential as a profiler for dimensional measurement has rarely been exploited because of its unreliable vertical accuracy. In NANO III, a STM was presented that was capable of 250 μm line scanning secured by a parallel spring mechanism in the X direction; it employed an interferometric monitoring STM tip, allowing feedback controlled motion in the Z direction. For this device, errors of step height and pitch measurements on a surface topography standard are estimated. The uncertainty of the 3 μm pitch and 100 nm step height measurements using the STM is 1.05 and 0.41 nm, respectively. The uncertainties are induced by thermal expansion, the motion of the scanners, the nonlinearity of the interferometer, and geometrical error. In the 100 nm, step height measurement, the sum of the standard deviation of the measured data and the total error is still in the subnanometer range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.