Abstract
Recent changes are updated for the proposed modification to use a fission plate converter and a matching filter/moderator assembly in a newly designed beam shutter to enhance the epithermal neutron beam for patient irradiations during boron neutron capture therapy (BNCT) at the Brookhaven Medical Research Reactor, Brookhaven National Laboratory. Engineering design of the new beam shutter system has been completed, and design review is under way. With this upgrade, the epithermal neutron flux could be increased up to seven times higher than the present beam, and the fast-neutron dose per epithermal neutron fluence could be reduced to near half. Such an enhanced epithermal neutron beam would increase the effectiveness of clinical BNCT by allowing increased doses to the tumor with similar or lesser doses to normal tissues.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.