Abstract

The concentration of glucose in biological fluids is in the micromolar range, the detection of which requires devices with high sensitivity and low limit of detection (LOD). Here, we report the real-time electronic detection of glucose using an antioxidant found in green tea, namely, epigallocatechin gallate (EGCG), decorated on carbon nanotubes (CNTs) and tested in a chemiresistor configuration. The detection principle relies on the spontaneous reaction of EGCG with hydrogen peroxide, a reactive oxygen species released during glucose oxidation, which is translated electrically as a change in CNT conductance. Our results suggest that the response of EGCG decorated CNTs was far superior to that of the bare CNT based device. The sensor detected glucose ranging from 10 nM to 1 μM with LOD of ∼8.7 nM, which is much lower than the commercially available finger-pricking based glucose sensors. This could pave the way for developing simple resistivity-based sensors capable of glucose detection in biological fluids other than blood, such as sweat and saliva.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.