Abstract

A simple and sensitive enzymatic method for determination of plasma and serum fatty acids (FAs) is described. The method is based on acylation of long chain FAs by a bacterial acyl-CoA synthetase (ACS) producing equivalent amounts of acyl-CoA and AMP. AMP production was measured using the coupled reaction of myokinase (MK), pyruvate kinase (PK) and lactate dehydrogenase (LDH) allowing fluorinate detection of NADH. Two moles of NAD were produced per mole of FA acylated. Concentrations of substrates and enzymes were kept as low as possible maintaining the ACS reaction as rate limiting. Addition of fat-free human serum albumin (HSA) to standards reduced initial reaction rates but did not affect end-point fluorescence levels. Triton X-100 partly counteracted the inhibition by HSA. To keep albumin concentration low, plasma or serum samples were diluted by 1:400. Duplicate measurements of plasma or serum FA concentrations between 0 and 2 mmol l-1 can then be performed on 5 microliters samples with intra- and inter-assay variation coefficients of 1.7 and 4% respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.