Abstract
Ion-selective electrode with cellulose triacetate membrane containing nonactin is employed for the potentiometric detection of ammonia produced in biocatalytic reaction in flow-injection system with enzyme reactor. The elimination of interferences occurring in the presence of alkali metal ions was achieved by covering a nonactin membrane with outer hydrophobic gas permeable membrane. The obtained flow-injection response to ammonia indicates a possibility of ammonia determination down to 10 μ M ammonia. In the flow-injection system for urea determination 200 μl of 10-fold diluted blood serum sample was injected into carrier stream of distilled water merged with TRIS buffer, passed through the urease flow-through reactor and then after merging with NaOH stream delivered to the detector. It was found in several series of natural blood serum samples, that the correction for endogenous ammonia in such a determination is not indispensable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.