Abstract

Two urea biosensors, based on carboxylated polyvinylchloride ammonium ion selective electrodes with urease immobilized with carbodiimide, were included in an array together with potassium and sodium ion selective electrodes and evaluated for simultaneous determination of urea and potassium in blood serum. The results show that it is possible to obtain principal component regression and partial least squares (PLS2) calibration models for simultaneous determination of urea (in the range 0.10–5 mM) and potassium (in the range 0.20–0.60 mM) based on a set of 9 calibration samples. The prediction errors of the concentration of urea and potassium in blood serum samples using this array and a PLS2 calibration model are lower than 13 and 5%, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.