Abstract

Liquid chromatography/electrospray ionization-tandem mass spectrometry (LC/ESI-MS/MS) enables the accurate and precise quantification of various analytes at very low concentrations, but it has room for improvement in analysis throughput. Multiplexing of samples in the same injection could be a promising procedure for enhancing the analysis throughput. This could be achieved by derivatization of the multiple samples with multiple isotopologous reagents. In this study, a sample-multiplexed LC/ESI-MS/MS assay using the 1-[(4-dimethylaminophenyl)carbonyl]piperazine (DAPPZ) isotopologues (2H0-, 2H3-, and 2H6-forms) was developed and validated for the simultaneous determination of primary bile acids in three different plasma samples in a single run. The developed method had satisfactory intra- and inter-assay precisions (≤ 2.3 and ≤ 4.2%, respectively) and accuracy (99.0 - 100.3%), and could reduce the total LC/ESI-MS/MS run time by more than 60% for 42 samples compared to the conventional method. Thus, the derivatization with the DAPPZ isotopologues worked well for enhancing the throughput of the LC/ESI-MS/MS assay of the bile acids.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.