Abstract

The effects of the germanium detector electronics on diffraction line patterns is investigated. It is shown that not only the detector resolution and the throughput but also the energy stability depend on both the specific detector settings and the dead time. For a moderate resolution versus throughput setting a correction function is proposed and applied to the near-surface residual stress analysis of three samples with considerably different stress states. It is demonstrated that without the correction function ghost stresses up to hundreds of MPa in the near-surface region are obtained. The correction procedure is verified by conventional X-ray measurements. In conclusion, the authors strongly suggest quantifying the electronic shifts of any individual detector systems prior to the analysis of residual stresses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.