Abstract

Eicosanoids, including prostaglandins and thromboxanes are lipid mediators synthetized from polyunsaturated fatty acids. They play an important role in cell signaling and are often reported as inflammatory markers. LC–MS/MS is the technique of choice for the analysis of these compounds, often in combination with advanced sample preparation techniques.Here we report a head to head comparison between an electrospray ionization source (ESI) and a new atmospheric pressure ionization source (UniSpray). The performance of both interfaces was evaluated in various matrices such as human plasma, pig colon and mouse colon. The UniSpray source shows an increase in method sensitivity up to a factor 5. Equivalent to better linearity and repeatability on various matrices as well as an increase in signal intensity were observed in comparison to ESI.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.