Abstract

The cochlear implant (CI) is the most successful neural prosthesis, restoring the sensation of sound in people with severe-to-profound hearing loss by electrically stimulating the cochlear nerve. Existing CIs have an external, visible unit, and an internal, surgically-placed unit. There are significant challenges associated with the external unit, as it has limited utility and CI users often report a social stigma associated with prosthesis visibility. A fully-implantable CI (FICI) would address these issues. However, the volume constraint imposed on the FICI requires less power consumption compared to today’s CI. Because neural stimulation by CI electrodes accounts for up to 90% of power consumption, reduction in stimulation power will result directly in CI power savings. To determine an energy-efficient waveform for cochlear nerve stimulation, we used a genetic algorithm approach, incorporating a computational model of a single mammalian myelinated cochlear nerve fiber coupled to a stimulator-electrode-tissue interface. The algorithm’s prediction was tested in vivo in human CI subjects. We find that implementation of a non-rectangular biphasic neural stimulation waveform may result in up to 25% charge savings and energy savings within the comfortable range of hearing for CI users. The alternative waveform may enable future development of a FICI.

Highlights

  • Sensorineural hearing loss (SNHL) is the most common sensory deficit and the most common congenital anomaly, affecting 360 million people worldwide[1]

  • Genetic algorithm (GA) simulations were performed for pulses having 25 μs cathodic and 25 μs anodic phases (50 μs total pulse duration)

  • The GA simulation was repeated for Phase widths (PW) of 50 μs and 100 μs in order to investigate the impact of PW on optimal GA shape (Fig. 1b)

Read more

Summary

Introduction

Sensorineural hearing loss (SNHL) is the most common sensory deficit and the most common congenital anomaly, affecting 360 million people worldwide[1]. The limited energy storage capacity of a FICI requires low-power (

Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call