Abstract
To enable an efficient and enduring decontamination of the rectal mucosa during transanal endosocopic procedures, we developed a device for reversible endolumenal colon occlusion (ColoShield). The aim of this study was to assess the value of ColoShield in reducing peritoneal contamination during a transrectal procedure. Sixteen pigs underwent transrectal hybrid NOTES cholecystectomy after standardized disinfective rectal washout either with endolumenal colon occlusion using ColoShield (N=8) or without colon occlusion (N=8). Rectal swab samples were taken before and after rectal washout and at the end of the procedure. Peritoneal biopsies for microbiological evaluation were obtained at the end of the procedure and at necropsy 7days after surgery. Peritoneal contamination at the end of surgery was significantly lower using ColoShield compared to not using colon occlusion [13 (1/8) vs. 75% (6/8); P=0.012]. No significant differences were found regarding contamination of rectal swabs and peritoneal contamination at necropsy. The application of ColoShield may increase the safety of transrectal NOTES and transanal endoscopic procedures by reducing peritoneal contamination and consecutive infectious complications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.