Abstract

BackgroundOur aim was to compare the efficacy of the end-tidal CO2-guided automated robot CPR (robot CPR) system with manual CPR and mechanical device CPR. MethodsWe developed the algorithm of the robot CPR system which automatically finds the optimal compression position under the guidance of end-tidal CO2 feedback in swine models of cardiac arrest. Then, 18 pigs after 11 min of cardiac arrest were randomly assigned to one of three groups, robot CPR, LUCAS CPR, and manual CPR groups (n = 6 each group). Return of spontaneous circulation (ROSC) and Neurological Deficit Score 48 h after ROSC were compared. ResultsA ROSC was achieved in 5 pigs, 4 pigs, and 3 pigs in the robot CPR, LUCAS CPR, and manual CPR groups, respectively (p = 0.47). Robot CPR showed a significant difference in Neurological Deficit Score 48 h after ROSC compared to manual CPR, whereas LUCAS CPR showed no significant difference over manual CPR. (p = 0.01; Robot versus Manual adjusted p = 0.04, Robot versus LUCAS adjusted p = 0.07, Manual versus LUCAS adjusted p = 1.00). ConclusionsThe end-tidal CO2-guided automated robot CPR system did not significantly improve ROSC rate in a swine model of cardiac arrest. However, robot CPR showed significant improvement of Neurological Deficit Score 48 h after ROSC compared to Manual CPR while LUCAS CPR showed no significant improvement compared to Manual CPR.