Abstract

E-6087 is a nonsteroidal anti-inflammatory compound that selectively inhibits cyclooxygenase-2. Because E-6087 has a chiral center, this compound is a racemic mixture of two stereoisomers, (+)-(R)-E-6087 (E-6231) and (-)-(S)-E-6087 (E-6232). A normal-phase liquid-chromatographic method for the enantioselective determination of E-6087 in human plasma was developed and validated. The samples were extracted using solid-phase extraction cartridges containing C(18) sorbent, and the extracts were redissolved in absolute ethanol and injected into the chromatographic system. The enantiomeric separation was achieved on a chiral stationary-phase column of derivatized amylose, and the enantiomers were quantified by fluorescence detection. The method was validated for drug concentrations ranging from 5 to 400 ng/ml for both enantiomers. No peaks interfering with the quantification of enantiomers were observed. The limit of quantification was 5 ng/ml, with precision expressed as a coefficient of variation lower than 10.6% and accuracy expressed as relative error lower than 12.2%. The utility of this method was demonstrated by analysis of plasma samples from healthy volunteers given an oral dose of rac-E-6087. Peak plasma levels of E-6231 were higher than levels obtained for E-6232. Results were consistent with those obtained with a conventional reversed-phase method used for determination of the racemic compound.