Abstract

A simple, specific and sensitive HPLC method has been developed for the determination of metoprolol and its two metabolites in human plasma and urine. Separation of metoprolol, α-hydroxymetoprolol, O-desmethylmetoprolol and esmolol (internal standard) was achieved on an Agilent XDB-C18 column (150mm×4.6mm, 5μm) using fluorescence detection with Ex 216nm and Em 312nm. The mobile phase consisted of ACN-H2O-0.1%TFA. The analysis was performed in less than 16min with a flow rate of 0.8mL/min. The assay was linear over the concentration range of 5-600ng/mL and 2.5-300ng/mL for metoprolol and its metabolites, respectively. The LOQ were 5.0 and 2.5ng/mL for plasma and urine, respectively. Good precision and accuracy for metoprolol and its two metabolites were obtained. The extraction recoveries were found to be more than 86.91% both in plasma and urine. At the same time, the method was successfully applied to nine healthy volunteers who had been given an oral tablet of 100mg metoprolol.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call