Abstract
X-ray fluorescence computed tomography (XFCT) is an emerging imaging modality that maps the three-dimensional distribution of elements, generally metals, in ex vivo specimens and potentially in living animals and humans. Building on our previous synchrotron-based work, we experimentally explored the use of a benchtop X-ray fluorescence computed tomography system for mapping trace-metal ions in biological samples. This system utilizes a scanning pencil-beam to stimulate the object and then relies on a detection system, with single or multiple slit apertures placed in front of position-sensitive X-ray detectors, to collect the fluorescence X-rays and to form 3-D elemental map without the need for tomographic imaging reconstruction. The technique was used to generate images of the elemental distributions of a triple-tube phantom and an osmium-stained zebrafish.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.