Abstract

An on-line flow injection preconcentration-ETAAS method is developed for trace determination of cobalt in biological materials and natural samples by ion-pair sorption on the inner walls of a PTFE knotted reactor. The ion-pair is formed between the negatively charged cobalt-nitroso-R-salt complex and the tetrabutylammonium counter-ion. An enhancement factor of 15, a sampling frequency of 17 and a concentration efficiency of 4 are obtained for a preconcentration time of 60 s and a sample loading flow rate of 5 mL min(-1). The detection limit (3sigma) is 5 ng L(-1). The relative standard deviation at the 0.2 microg L(-1) level is 2.3%. The analytical results obtained for standard reference materials are in good agreement with the certified or indicated values and satisfactory recoveries of spiked cobalt in tap water are obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.