Abstract

Voltage Sweep Ion Mobility Spectrometry (VSIMS) has been applied to complex samples using electrospray ionization (ESI). The usable range of VSIMS has been extended from that obtained in previous studies where only volatile compounds were investigated. Using ESI, VSIMS was evaluated with compounds with reduced mobility values as low as 0.3 V2cm−1 s−1. The primary advantage of VSIMS is to enable a drift time ion mobility spectrometer (DTIMS) to detect both fast and slow moving ions at optimal resolving power, thus improving the peak capacity. In this work ESI-VSIMS was applied to a series of small peptides and drugs spanning a large range of reduced mobility values in order to demonstrate ESI-VSIMS to separation. To demonstrate improved peak capacity of IMS with voltage scan operation, oligomers of silicone oil provided a series of evenly-spaced peaks, ranging in reduced mobility values from 0.85 to 0.3 V2cm−1 s−1. The peak capacity of 61 for a standard IMS was improved to 102 when voltage sweep operation was employed. In addition, VSIMS increased the average resolving power of the DTIMS from 66 to 106 for silicone oil.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.