Abstract
To improve the sensitivity of the UV-detection for the determination of trace amounts of albumin by capillary zone electrophoresis (CZE), five on-line preconcentration techniques, including field-amplified sample stacking (FASS), head-column field-amplified sample stacking (HC-FASS), stacking with a polymer solution, dynamic pH junction and large volume sample stacking (LVSS ) with reversed polarity, were compared. Sensitivity enhancement factor and reproducibility were two factors that were used to assess the suitability of each method. To minimize protein adsorption on the capillary wall, capillaries were covalently modified with anionic polymer, poly(sulfopropylmethacrylate) coating. All used methods have good reproducibility. The maximum sensitivity enhancement factor (about 67-fold in terms of peak heights) was achieved with LVSS technique. The concentration limit of detection (LOD) (S/N = 3) for the human serum albumin obtained with the optimized LVSS approach was 15 μg/ml with UV-detection. The method was further evaluated for the analysis of urine samples with gel-filtration-based sample-desalting procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.