Abstract
A wide holographic field of view (up to 15 μm in the Hitachi-HF2000) is achieved in a TEM by switching off the objective lens and imaging the sample by the first intermediate lens. Fig.1 shows the corresponding ray diagram for low magnification image plane off-axis holography. A coherent electron beam modulated by the sample in its amplitude and its phase is superimposed on a plane reference wave by a negatively biased Möllenstedt-type biprism.Our holograms are acquired utilizing a Hitachi HF-2000 field emission electron microscope at 200 kV. Essential for holography are a field emission gun and an electron biprism. At low magnification, the excitation of each lens must be appropriately adjusted by the free lens control mode of the microscope. The holograms are acquired by a 1024 by 1024 slow-scan CCD-camera and processed by the “Holoworks” software. The hologram fringes indicate positively and negatively charged areas in a sample by the direction of the fringe bending (Fig.2).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings, annual meeting, Electron Microscopy Society of America
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.