Abstract

AbstractA significant loss in electron probe current can occur before the electron beam enters the specimen chamber of an environmental scanning electron microscope (ESEM). This loss results from electron scattering in a gaseous jet formed inside and downstream (above) the pressure-limiting aperture (PLA), which separates the high-pressure and high-vacuum regions of the microscope. The electron beam loss above the PLA has been calculated for three different ESEMs, each with a different PLA geometry: an ElectroScan E3, a Philips XL30 ESEM, and a prototype instrument. The mass thickness of gas above the PLA in each case has been determined by Monte Carlo simulation of the gas density variation in the gas jet. It has been found that the PLA configurations used in the commercial instruments produce considerable loss in the electron probe current that dramatically degrades their performance at high chamber pressure and low accelerating voltage. These detrimental effects are minimized in the prototype instrument, which has an optimized thin-foil PLA design.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.