Abstract

Migration of a cochlear implant electrode is a hitherto uncommon complication. So far, array migration has only been observed in lateral wall electrodes. Between 1999 and 2014, a total of 27 patients received bilateral perimodiolar electrode arrays at our institution. The insertion depth angle was estimated on the initial postoperative scans and compared with the insertion depth angle of the postoperative scans performed after contralateral cochlear implantation. Seven (25.93%) patients were found to have an electrode array migration of more than 15°. Electrode migration in perimodiolar electrodes seems to be less frequent and to occur to a lower extent than in lateral wall electrodes. Electrode migration was clinically asymptomatic in all cases.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.