Abstract

Objective Electrode impedances play an important role in cochlear implant patient management. During clinical visits, electrode impedances are calculated from a single point voltage waveform. In the present study, multipoint electrode impedance analysis was performed to study electrode impedance and its subcomponents in patients with three different types of cochlear implant electrode arrays. Design Voltage waveforms were measured at six different time points during the cathodic phase of a biphasic pulse in forty-seven cochlear implant patients with perimodiolar, mid-scala, or lateral wall electrode arrays. Multipoint electrode impedances were used to determine access resistance and polarization impedance. Results Access resistance of approximately 5 kΩ was calculated across the three different electrode arrays. Mid-scala electrodes showed a smaller increase in impedances as a function of pulse duration compared to the other electrodes. Patients with lower impedances showed higher capacitance and lower resistance, suggesting that differences in electrochemical reaction at the electrodes’ surface can influence impedances in cochlear implants. Conclusions Analysis of cochlear implant electrode impedances and their subcomponents provides valuable information about resistance to the flow of current between stimulating and return electrodes, and build an understanding of the contribution of electrochemical processes used to deliver electrical stimulation to the auditory nerve.