Abstract

Novel electrochemical methods have been developed for determination of total hemoglobin, hematocrit, and detection of hemolysis in whole blood. Hemoglobin is measured through its peroxidase activity, a fluoride ion-selective electrode being used to monitor the rate of fluoride ion production from the oxidation of an organofluorine compound. Results agree well with those obtained with the cyanmethemoglobin method (r = 0.970). Hematocrit is determined from the ratio of the sodium ion concentrations measured with an ion-selective electrode before and after lysis of the erythrocytes. Results by this and the microhematocrit method correlated well (r = 0.987). Hemolysis in a whole-blood sample is detected by using an oxygen electrode to measure the oxygen released when hemoglobin in plasma is oxidized.