Abstract

A flow-through two electrode wall-jet cell with a platinum measuring electrode and a cell volume of 20 nl has been designed and evaluated. It has been used to detect phenols by reversed phase liquid chromatography using short micro bore columns. The linear dynamic range between the measured current and the concentration is greater than 103 (1.5×10−7−5×10−4 mol/l) and the minimum analyzable amount was found to be 10 pg for pyrocatechol. A negligible broadening in the detector permits the use of micro columns down to 0.5 mm internal diameter, packed with 5 μm particles, without any substantial distortion of eluted zones.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.