Abstract
A method for the detection of ascorbic acid (AA) was developed using square wave anodic stripping voltammetric (SWASV) analysis with a glassy carbon electrode (GCE). The results indicated a sensitive oxidation peak current of AA on the GCE. A linear curve was obtained within a concentration range of 1-130.0 microg/L with a preconcentration time of 270 s. The relative standard deviation of 30.0 microg/L observed was 0.1016% (n = 12) under optimum conditions. The low detection limit (S/N) was pegged at 0.30 microg/L. Results showed that the method developed can be used to assay biological and pharmaceutical samples, and food samples, as well as other materials requiring AA analyses.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.