Abstract

The fabrication of silver (Ag) and gold (Au) thin film electrodes was successfully carried out using the DC sputtering deposition method. These thin film electrodes were able to detect the increase in serum albumin concentration that was used as a prognostic factor for leukemia. The simulation and the optical experimental analysis show that an increase in BSA concentration can increase the absorbance peak observed at a wavelength of 435 nm on hypoalbumin medium and 470 nm on normal concentration of serum albumin medium. The performance of the electrodes was electrochemically tested, in which it was shown that a decrease in oxidation and reduction peaks occurred with respect to an increase in BSA concentration. An oxidation peak was observed at a voltage of 0.5 V for the Ag thin film. For the Au, Au/Ag, and Ag/Au thin films, an oxidation peak was observed at a voltage of 1.0 V. The limits of detection (LODs) of the Ag, Ag/Au, Au, and Au/Ag thin films were 0.56, 0.24, 0.64, and 0.36 g/dL, respectively. Therefore, based on both the electrochemical and optical analysis, the Ag/Au thin film possessed the highest potential for prognosis monitoring of leukemia compared with the other Ag and Au thin films.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call